![Mettler Electronics Sys*Stim 294 Instruction Manual Download Page 40](http://html1.mh-extra.com/html/mettler-electronics/sys-stim-294/sys-stim-294_instruction-manual_1780253040.webp)
Mettler Electronics Corp.—
Rev.F_09/13/17
40
High Volt Mode
75 us
0 volts
500 volts
50 %
8 us
Figure 8.5—High Volt
Waveform
Waveform Type:
Monophasic twin peak
Polarity:
Positive
or
negative
Volts:
500 volts peak, 1 Kohm load
Current:
0 –500 mA peak,
1 Kohm load
Average current at
maximum intensity
and frequency:
1.2 mA at 120 pps with
1
Kohm
load
Maximum current
density under 2"
diameter electrode.
0.06 mA/cm²
Frequency:
1–120
Hz
Frequency Modulation:
No
Phase Duration:
8
s at 50% Vmax
Polarity:
Positive
or
negative
Available Amplitude
Modulation Options:
Continuous
Surge
Available Channels:
Channel One only
Microcurrent Mode
+
-
or
+
-
Figure 8.6—Microcurrent
Waveform
Waveform Type:
Monophasic or biphasic
square
wave
Polarity:
Positive or negative or
biphasic
pulses
Volts:
1 Volt peak, 1 Kohm load
Current:
10-990
A peak, 1 Kohm load
Average current at
maximum intensity
and frequency:
990
A
Maximum current
density under 2"
diameter electrode.
24.4
A/cm²
Frequency:
0.5-500
Hz
Duty
Cycle:
50%
Frequency Modulation:
No
Pulse Duration:
1-1000 ms
Available Amplitude
Modulation Options:
Continuous
Available Channels: