Mettler Electronics Corp.—
Rev.F_09/13/17
14
Electrode Cable Connections
Figure 3.2— Sys*Stim 294, Front View—
Electrode Cable Connections
3.2 EMC Guidance
CAUTION:
Medical Electrical Equipment needs special precautions regarding Electromagnetic
Compatibility (EMC) and needs to be installed and put into service according to the EMC
information provided in the following tables.
Portable and mobile Radio Frequency (RF) communications equipment can affect Medical
Electrical Equipment.
Accessories:
Hospital Medical grade power cord of a maximum length of 120 inches
WARNING:
The use of accessories, other than those specified, except those supplied or sold by Mettler
Electronics Corp., as replacement parts for internal or external components, may result in
increased EMISSIONS or decreased IMMUNITY of the Sys*Stim 294.
Guidance and manufacturer’s declaration – electromagnetic emissions
The Sys*Stim 294 is intended for use in the electromagnetic environment specified below. The customer or
the user of the Sys*Stim 294 should assure it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The Sys*Stim 294 must emit electromagnetic energy in order to
perform its intended function. Nearby electronic equipment
may be effected.
RF emissions
CISPR 11
Class B
The Sys*Stim 294 is suitable for use in all establishments other
than domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Applicable
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Applicable