![Mettler Electronics Sys*Stim 294 Instruction Manual Download Page 23](http://html1.mh-extra.com/html/mettler-electronics/sys-stim-294/sys-stim-294_instruction-manual_1780253023.webp)
Sys*Stim 294 Instruction Manual—
Rev.F_09/13/17
23
Interferential—
Continuous
Premodulated—
Continuous,
Surge and Reciprocation
Medium Frequency,
Russian
—
Continuous,
Surge and
Reciprocation
Biphasic—
Continuous,
Surge and Reciprocation
High Volt—
Continuous and Surge
Microcurrent—
Continuous
1 0 1 0
min s
s
s
Hz
Surge Mode—
Press Surge selector until you see the On/Off duty cycle that
you would like to use. Press the
key to accept the values.
Preset On/Off choices are 10 10, 10 20, 10 30, 10 40,
10 50 and 10 60.
If you press the Surge selector one more time after “10 60” is
displayed, a single number is displayed. This represents the
last On time that was programmed. To change the number,
use the numeric keypad to enter a new value followed by the
key. The Off time is then displayed, enter a new value
followed by the
key.
1 0
min s
s
s
Hz
Reciprocation Mode—
To setup a Reciprocation program you must have stimulation
setup for two-channel operation. If Channel 1 is lit, press
Channel 2. If Channel 3 is lit, press Channel 4.
Press the Reciprocation key. Enter a value from 2 to 240 and
press the
key.
Surge + Reciprocation—
Press Reciprocation key. Press the
key. Press the Surge
key until you see the On/Off time you want to use. Press the
key.
1 5. 0 0
min s
s
s
Hz
1
2 3
4 5 6
7 8 9
0
6.
Enter the treatment time using the numeric keypad. The
maximum treatment time is 60 minutes. If you do not enter a
time, the time will count up during a treatment session, but
will not exceed 60 minutes.
7.
Apply the electrodes to the patient. Attach the electrode
cables to the electrodes.