
Sys*Stim 240 Instruction Manual — Rev.I_09/23/19
55
9.
Error E80 - Overcurrent Error
Appears when the current level spikes (e.g., the
electrode is partly detached from the treatment area
or moved out of position).
Press “Enter” to return to the waveform screen. Check
the electrodes. Check the electrode cable connections
to make sure that they are connected. Make sure both
electrodes are attached to the cables and to the
patient. Try fresh electrodes and or cables to resolve
this problem.
Press the “Start/Pause” key and to restart the
treatment and return to the previously set intensity.
Or press the “Stop” button to stop the treatment and
return to the treatment screen. Pressing the
“Start/Pause” key will restart the treatment at the
beginning. You will need to press “Enter” to begin to
increase the intensity and start the treatment.
If the above does not correct the error. Remove the
electrode cables from the unit. Then turn the unit Off
and turn it back ON and retry the treatment. If this
does not correct the problem, the unit requires
servicing. Report the error to service provider to assist
with hardware troubleshooting.
10.
Error E85 – Current too low
Check the electrodes. Resume the treatment by doing
the following:
Press “Enter” to return to the waveform screen. Check
the electrodes. Check the electrode cable connections
to make sure that they are connected. Make sure both
electrodes are attached to the cables and to the
patient. Try fresh electrodes and or cables to resolve
this problem.
Press the “Start/Pause” key and to restart the
treatment and return to the previously set intensity.
Or press the “Stop” button to stop the treatment and
return to the treatment screen. Pressing the
“Start/Pause” key will restart the treatment at the
beginning. You will need to press “Enter” to begin to
increase the intensity and start the treatment.
If the above does not correct the error. Remove the
electrode cables from the unit. Then turn the unit Off
and turn it back ON and retry the treatment. If this
does not correct the problem, the unit requires
servicing. Report the error to service provider to assist
with hardware troubleshooting.