Mettler Electronics Corp. — Rev.I_09/23/19
22
6.
Use the M Wheel to scroll down to the Language you want for all of
the menus and screens and press enter.
7.
The new language will be set and you will return to the Information
Menu.
3.4 EMC
Guidance
CAUTION:
Medical Electrical Equipment needs special precautions regarding Electromagnetic
Compatibility (EMC) and needs to be installed and put into service according to the
EMC information provided in the following tables.
Portable and mobile Radio Frequency (RF) communications equipment can affect
Medical Electrical Equipment.
Accessories:
Hospital Medical grade power cord of a maximum length of 120 inches or 3 meters
WARNING:
The use of accessories, other than those specified, except those supplied or sold by
Mettler Electronics Corp., Incorporated as replacement parts for internal or
external components, may result in increased EMISSIONS or decreased IMMUNITY
of the Sys*Stim 240.
Guidance and manufacturer’s declaration – electromagnetic emissions
The Sys*Stim 240 is intended for use in the electromagnetic environment specified below. The
customer or the user of the Sys*Stim 240 should assure it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The Sys*Stim 240 must emit electromagnetic energy in
order to perform its intended function. Nearby
electronic equipment may be effected.
RF emissions
CISPR 11
Class A
The Sys*Stim 240 is suitable for use in all establishments
other than domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Applicable
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Applicable