Sys*Stim 240 Instruction Manual — Rev.I_09/23/19
35
4.8 Premodulated (IFC, 2-Pole) Procedure
1.
Turn on the mains power switch located on the back of the unit by
pressing dot on the switch to the “On” position. Image shows
switch in “Off” position.
2.
Press either “1” or “2” to select channel one or two. Please note
that Reciprocation requires 2 channels.
3.
Insert the electrode cable into the channel connector for the
channel that you will be using.
4.
Place the electrodes on the patient.
5.
Insert the patient safety cable into the jack located on the back of
the unit and hand the end with the button to the patient letting
him/her know that they can push this button at any time during
the treatment to stop it.
6.
Use the M Wheel to scroll to any of the parameters that you would
like to change. The parameters and their ranges are listed here for
your reference:
Carrier frequency:
2500, 4000 or 5000 Hz
Interference frequency: 1-250 Hz
Frequency Modulation: Low set: 1-250 Hz
High
set:
1-250
Hz
Preset Frequency Sweeps: 1-15 Hz, 80-150 Hz, 1-150 Hz
Amplitude Modulation:
Surge:
On (s)/Off (s) 5/5, 4/12, 10/10, 10/20, 10/30,
10/50,
Manual:
1-240/1-240
Recip: Ch1 (s)/Ch2 (s) 5/5, 4/12, 10/10, 10/20, 10/30,
10/50,
Manual:
1-240/1-240
Ramp:
0.5, 1, 2 or 5 seconds
Type:
CC
or
CV
Options:
Channel(s) 1 or 2
7.
If all of the parameters are set correctly, position the cursor over
the output window and press “Enter”. Then use the M Wheel by
dragging your finger in a clockwise direction to adjust the intensity
up. You will notice that if two channels are being used, they will go
up simultaneously. The intensity in the digital display will go up
and the green bar will get larger as the intensity increases. Moving
your finger around the M Wheel in a counterclockwise direction
will decrease the intensity. The time will begin to count down until