Sys*Stim 240 Instruction Manual — Rev.I_09/23/19
17
Section 3—Installation
3.1 Installation
Instructions
1.
When installing the unit, pay attention to the following:
Install the unit beyond the reach of possible water splashes.
Install the unit where it will not be adversely affected by atmospheric pressure,
temperature, humidity, sunlight, dust, ventilation, salt air, sulfur, or other such harmful
substances.
Protect the unit against instability, vibration, or impact (including during transportation).
Do not leave the unit in locations with combustible airborne materials such as combustible
anesthetic gases mixed with oxygen, nitrogen suboxide and air, or combustible disinfecting
agents or cleaning agents mixed with air.
Do not install the unit where chemical products are stored or where gases may be emitted.
2.
The Sys*Stim 240 may be susceptible to interference originating from shortwave diathermy
units operating in close proximity to it. Avoid operating the Sys*Stim 240 adjacent to and
simultaneously with operating shortwave devices.
3.
If you have chosen the optional battery pack, install the battery as seen in Figures 3.1 and 3.2.
4.
To install the optional battery first remove the power cord. Then, remove the screws located
on the back panel of the Sys*Stim 240 on either side of the specification label using a Phillips
screw driver, Figure 3.1.
5.
Place the battery into the compartment as shown in Figure 3.2. Turn the unit upside down
while installing the battery.
6.
Close the battery compartment lid and reattach it by tightening the Phillips head screws.
7.
Connect the line cord to the back of the Sys*Stim 240. (See Figure 3.3)
8.
Plug the line cord into a grounded wall outlet that is rated at 100-240 VAC 50/60 Hz. Your
mains power supply must match the voltage requirements listed on the serial number label of
your device.
Do not connect the Sys*Stim 240 to a power supply rated differently than that
described above.
9.
The line cord comes equipped with a standard 3–prong plug. This plug provides grounding for
the Sys*Stim 240. Do not defeat its purpose by using 3–to–2 prong adapters or any other
means of attaching to a wall outlet.
10.
If the optional battery is installed, it will begin charging as soon as the line cord is plugged into
the wall and the Sys*Stim 240 is turned on using the switch on the back of the unit. The
charging status will be displayed on the display. Full charging takes up to four hours. The unit
may be used while it is charging.
11.
Plug the electrode cables (ME 2260) into the electrode cable connections as seen in Figure 3.4.
12.
If you have chosen to purchase either the optional laser or cluster kit, then install the
applicator holders in Figure 3.5 on the side of the unit using the Phillips head screws included
with the holders. Plug in the applicator into the front of the device as seen in Figure 3.6 and
insert into the holder on the side of the unit. When using either the cluster or the laser