![METRON QA-ST User And Service Manual Download Page 8](http://html1.mh-extra.com/html/metron/qa-st/qa-st_user-and-service-manual_1779892008.webp)
Resolution:
1 V L-N
2.
Current Capacity
15 Amps @ 100 Volts
15 Amps @ 120 Volts
10 Amps @ 230 Volts
3.
Resistance Measurement
DC-Source:
Range:
0 - 19.99 Ohms
Accuracy:
± 2%reading, ± 1 digit
Resolution:
0.01 ohms
Current Source:
100 mA DC
AC-Source:
Range:
0 - 1999 mOhms
Accuracy:
± 2% reading, ± 1 digit
Resolution:
1 mOhm
Current Source:
200 mA AC
4.
Leakage Current
Range:
0 - 1999 µA RMS
Accuracy:
DC and 25 to 1.0 kHz, ± 2% reading, ± 1 digit
1.0 kHz to 100 kHz, ± 2.5%R, ± 1 digit
100 kHz to 1 MHz, ± 5.0%R, ± 1 digit
Resolution:
1 µA
5.
Input Impedance
1000 ohm, AAMI ES1-1993 load or
IEC 60601.1 load
6.
Power Consumption
110 V (50/60) Hz = 200 mA
115 V (50/60) Hz = 200 mA
240 V (50/60) Hz = 100 mA
2