6.
Press “Patient leakage current lead to GND” (Ground).
7.
Adjust the voltage source until the ammeter displays the first
reference value in the test protocol. QA-ST will display the
same value as the reference ammeter. Note the value.
8.
Perform the test for all the given values in the test protocol
under “Patient Leakage Current Lead to GND: Reference”.
12