![METRON QA-ST User And Service Manual Download Page 36](http://html1.mh-extra.com/html/metron/qa-st/qa-st_user-and-service-manual_1779892036.webp)
10. Perform the test for tall the given values in the test protocol
under “Leakage Current Int. Case: Reference”.
5.5 Le
akage
Cur-
rent –
Exter-
nal
Case
The purpose of this test is to determine if there is any leakage cur-
rent between DUT Enclosure (chassis) and Mains Earth.
6.
Connect a voltage source in series with an ammeter and
10K
Ω
resistance as shown below.
7.
Press “Leakage current Ext. Case”.
8.
Adjust the voltage until the ammeter displays the first refer-
ence value in the test protocol. QA-ST will display the same
value as the reference ammeter. Note the value.
9.
Next, press “Outlet Supply RS” (Reversed Supply). The dis-
played QA-ST value will not change.
7
Summary of Contents for QA-ST
Page 1: ...QA ST User and Service Manual QA ST Safety Tester North American Version 1 1...
Page 4: ...This page intentionally left blank iv...
Page 6: ...This page intentionally left blank vi...
Page 13: ...This page intentionally left blank 3...
Page 20: ...4...
Page 29: ...13...
Page 44: ...Appendix A Circuit Board Schematic North American Version A 1 A 1...