6.
Press “Patient leakage current – lead to lead”.
7.
Adjust the voltage until the ammeter displays the first reference
value in the test protocol. QA-ST will display the same value as
the reference ammeter. Note the value.
8.
Press “lead to lead” to change the patient lead number. The
diode panel will display, in binary format, the number of the
actual patient lead that is under test.
9.
Perform the test for tall the given values in the test protocol
under “Patient Leakage Current Lead to Lead: Reference”.
10