![METRON QA-ST User And Service Manual Download Page 32](http://html1.mh-extra.com/html/metron/qa-st/qa-st_user-and-service-manual_1779892032.webp)
5
.
2
M
a
i
n
s
S
u
p
p
l
y
V
o
l
t
a
g
e
The purpose of this test is to calibrate the QA-ST to display the cor-
rect mains voltage.
3. Press
“Supply V”.
4. Adjust the mains voltage with an external variable AC voltage
(VARIAC) source. Then, adjust the QA-ST potentiometer R67
(see Appendix A for location)
until the QA-ST LCD displays the
same value.
5. Perform the test for tall the given values in the test protocol under
“Mains Supply Voltage: Reference”.
3
Summary of Contents for QA-ST
Page 1: ...QA ST User and Service Manual QA ST Safety Tester North American Version 1 1...
Page 4: ...This page intentionally left blank iv...
Page 6: ...This page intentionally left blank vi...
Page 13: ...This page intentionally left blank 3...
Page 20: ...4...
Page 29: ...13...
Page 44: ...Appendix A Circuit Board Schematic North American Version A 1 A 1...