the User Offset Zero file has been corrupted;
•
the ADC driver returned an unknown error code;
•
the command parser or dispatcher returned an unknown error code;
•
encountered an error formatting a result;
•
could not open one of the data files file;
•
unable to perform reset;
•
encountered an unknown choice for character data;
•
encountered something other than a value, MINimum, MAXimum or
DEFault for a numeric parameter;
•
encountered something other than a Boolean or DEFault for a Boolean
parameter; or
•
encountered an unknown parameter type.
Please report to Metrolab the circumstances under which you received this error.
8-6-2
204: Data buffer was overrun
In timed-trigger mode, data was lost because the read-out did not keep up with
the acquisition.
8-6-3
205: Measurements were over-range
The field value exceeded the selected range.
8-6-4
206: Timer was overrun
In timed trigger mode, a measurement was still in progress when the timer to start
the next measurement expired. The next measurement was skipped.
8-6-5
207: Bad data compression
When returning data in PACKed FORMat, the true dynamic range of the data
could not be accurately represented by the delta values. The maximum delta
value was used, while attempting to correct the error on the subsequent
sample(s). It is recommended to use a lesser degree of compression.