THM1176/TFM1186 Use
r’s Manual
v 2.0 r 1.0
– 02/20
Copyright © 2020 Metrolab Technology SA
www.metrolab.com
36
Command
Parameters
Description
:READ
Abort pending triggers and
perform a measurement
with existing parameters.
Note: cannot be used when
TRIGger:SOURce is BUS.
[:SCALar]
Perform a single
measurement. Set the range
for the expected value if
provided; use the previously
selected range if not. Return
at least the requested
number of significant digits.
[:FLUX]
<expected_value>
min=0T, max=20T, def=0T
<digits>
min=1, max=5, def=3
:X?
[<expected_value>]
[,<digits>]
Return x-component of flux
density measurement
[:Y]?
[<expected_value>]
[,<digits>]
Return y-component of flux
density measurement
:Z?
[<expected_value>]
[,<digits>]
Return z-component of flux
density measurement
:ARRay
Perform a series of <size>
measurements. The other
parameters are as for
:READ:SCALar.
[:FLUX]
<size>
min=1, max=2048, def=1
<expected_value>
min=0T, max=20T, def=0T
<digits>
min=1, max=5, def=3
:X?
<size>
[,[<expected_value>]
[,<digits>]]
Return x-component of flux
density measurement
[:Y]?
<size>
[,[<expected_value>]
[,<digits>]]
Return y-component of flux
density measurement
:Z?
<size>
[,[<expected_value>]
[,<digits>]]
Return z-component of flux
density measurement