-225: Out of memory
The device has insufficient memory to perform the requested operation:
•
when writing the new User Offset Zero file to Flash.
8-4 -400: QUERY ERRORS
8-4-1
-400: Query error
Generic query error:
•
Device being cleared when query was received.
On the THM1176/TFM1186, no other conditions should generate this error. If this
is not the case, please report to Metrolab the circumstances under which you
received this error.
8-4-2
-410: Query INTERRUPTED
The host has sent a new command before finishing reading the response to a
preceding query.
8-4-3
-420: Query UNTERMINATED
The host is trying to read a response without having sent a complete query.
8-4-4
-440: Query UNTERMINATED after indefinite response
Indicates that a query was received in the same program message after a query
requesting an indefinite response was executed. On the THM1176/TFM1186, the
only command returning an indefinite response (“Arbitrary ASCII Response Data”)
is *IDN.
8-5 100: INSTRUMENT-DEPENDENT COMMAND ERRORS
8-5-1
101: Invalid value in list
One or more values in a numeric list parameter are invalid, e.g. floating point
when not allowed.
8-5-2
103: Wrong units for parameter
A parameter within the command has the wrong type of units for the command.
8-6 200: INSTRUMENT-DEPENDENT EXECUTION ERRORS
8-6-1
200: Software Error
The firmware has encountered an unexpected error: