THM1176/TFM1186 Use
r’s Manual
v 2.0 r 1.0
– 02/20
Copyright © 2020 Metrolab Technology SA
www.metrolab.com
29
Magnetic field strength
T
M, U, N
Tesla (default)
MAHZP
Megahertz proton
GAUSS
K, M
Gauss
Other
S
M, U
Seconds
The following tables list the legal commands for the THM1176 family, in
alphabetical order.
Command
Parameters
Description
:ABORt
Reset the trigger system:
- Aborts acquisition in
progress
- Disables trigger
- Disables continuous trigger
Command
Parameters
Description
[:CALCulate]
:AVERage
:COUNt?
[MINimum | MAXimum |
DEFault]
Query averaging count
:COUNt
<count> | MINimum |
MAXimum | DEFault
Set averaging count
Command
Parameters
Description
:CALibration
[:INITiate]
Initiate the offset correction
procedure in zero-gauss
chamber
:ZERO
Restore factory offset
correction
:STATe?
[DEFault]
Query calibration state
:STATe
<boolean> | DEFault
Set calibration state:
whether or not temperature
and gain calibration is
applied. ON by default.