Command
Parameters
Description
:MEASure
Abort any pending triggers
and perform measurements
using the default acquisition
parameters:
- Immediate trigger
- Continuous initiation off
- Averaging count 1
- Calibration state on
- Default range
[:SCALar]
Perform a single
measurement. Set the range
for the expected value if
provided; auto-range if not.
Return at least the
requested number of
significant digits.
[:FLUX]
<expected_value>
min=0T, max=20T, def=0T
<digits>
min=1, max=5, def=3
:X?
[<expected_value>]
[,<digits>]
Return x-component of flux
density measurement
[:Y]?
[<expected_value>]
[,<digits>]
Return y-component of flux
density measurement
:Z?
[<expected_value>]
[,<digits>]
Return z-component of flux
density measurement
:ARRay
Perform a series of <size>
measurements. The other
parameters are as for
:MEASure:SCALar.
[:FLUX]
<size>
min=1, max=2048, def=1
<expected_value>
min=0T, max=20T, def=0T
<digits>
min=1, max=5, def=3
:X?
<size>
[,[<expected_value>]
[,<digits>]]
Return x-component of flux
density measurement
[:Y]?
<size>
[,[<expected_value>]
[,<digits>]]
Return y-component of flux
density measurement
:Z?
<size>
[,[<expected_value>]
[,<digits>]]
Return z-component of flux