4.1 Sample with air gap
■■■■■■■■■■■■■■■■■■■■■■
24
■■■■■■■■
863 Compact Autosampler
4 Automation sequences
NOTICE
The following method sequences and parameters can only be used if
the 863 Compact Autosampler is
not
controlled by a PC software (e.g.
MagIC Net) via USB connection.
4.1
Sample with air gap
This method template is suitable for simple determinations.
An air gap is formed in the transfer tubing between the individual sam-
ples.
Preparing and starting the sample series
1
Press
[START]
.
2
Under
Number of samples
, enter the number of samples or the
maximum value of
99
.
3
Under
Next sample pos.
, enter the rack position of the first sample
to be processed.
4
Press
[BACK]
.
5
Start the sample series in the
IC Net
software.
NOTICE
The method run will be automatically restarted for every sample via
remote connection (Autostart). A stepping pulse of the connected
instrument is expected at the end of the method run. The 863 Compact
Autosampler must be stopped manually by means of the button
[STOP]
after the processing of the last sample.
The individual steps of the method:
■
Move to the sample