3 Installation
863 Compact Autosampler
■■■■■■■■
15
3 Inserting the needle holder
Insert the needle holder in the lift head and screw it tight with the
nut from below.
4 Inserting the needle
■
Loosen the pressure screw from the needle holder and remove it.
■
Insert the needle from above partway into the opening of the
needle holder.
■
Slide the PEEK ferrule down over the needle from above. The nar-
row side of the seal must face upwards.
5 Fastening the needle
■
Screw the pressure screw into the needle holder from above.
Lightly push the needle upwards from below during the process.
■
Tighten the pressure screw firmly.
6 Connecting the capillary
■
Slide a 6.2744.070 PEEK pressure screw over the end of the capil-
lary.
■
Manually tighten the PEEK pressure screw with the capillary on
the nut of the needle holder. The capillary must be pushed in
while doing so.
7 Closing the safety shield
Tip down the safety shield and fix it with the fastening screws.
WARNING
The safety shield must always be closed for safety reasons during opera-
tion of the 863 Compact Autosampler.