1.2 Development of a method
799 GPT Titrino: Getting started
6
Press <PRINT> <REPORTS> <ENTER> to print your re-
ports. Your printout will look as follows:
Identification of report type (cu = curve)
Start volume and method identification
Scaling of curve axis
Instrument type with identification and pro-
gram version
Initial pH
Volume and pH value of EP1
Calculated result
Entry of a stop criterion
<PARAM>
<
↓
>
<ENTER>
Entries <ENTER>
2 x <QUIT>
You don't want to stop your titration manually each time.
Let us enter an automatic stop criterion.
Press <PARAM> and place the cursor to ">stop condi-
tions". Press <ENTER> to go to this inquiry group.
The absolute stop volume. It can be used as a safety
volume to prevent overflow of the titration vessel.
Enter a value for the stop volume that appears suitable
for your titration vessel.
Set the pH value, e.g. 11.5 as stop criterion.
If several stop criteria have been set, the one that is
reached first, applies.
Confirm the input with <ENTER> and quit the inquiry
with <QUIT>.
Try another titration with this method.
parameters
>stop conditions
stop
V:
abs.
stop V 99.99 ml
stop pH aus
stop EP 9
filling rate max. ml/min
'cu
799 GPT Titrino 01102 799.0010
date 2002-04-08 time 09:07 1
Start V 0.000 ml DET pH ********
1.0 ml/div dpH=2.0/div
−−−−−−−−−−−−−−−−
'fr
799 GPT Titrino 01102 799.0010
date 2002-04-08 time 09:07 1
pH(init) 2.88 DET pH ********
smpl size 2 ml
EP1 2.083 ml 7.64
RS1 3.80 g/l
manual stop
−−−−−−−−−−−−−−−−