2.3
Mode SET, parameters
799 GPT Titrino: Titration, dosing and measuring modes
15
2.3 Mode SET, parameters
Display Meaning
Input
range
>SET1
EP at pH
dynamics
max.rate
min.rate
stop crit:
stop drift
t(delay)
stop time
Individual parameters for EP1.
Preset EP1 at pH, U, or I, resp.
Distance from EP where constant dosing
stops and controlling begins.
Maximum dosing rate.
Minimum dosing rate.
Type of stop criteria.
Titration stops if stop drift is reached.
Stop if there is no dosing during t(delay).
If t(delay) is ”INF”:
stop after a time.
pH: 0.00...±20.00,
OFF
(0...±2000 mV, 0.0...±200.0 µA)
pH: 0.01...20.00,
OFF
(1...2000 mV, 0.1...200.0 µA)
0.01...
10
...150 mL/min,
max.
0.01...
25.0
...9999 µL/min
drift
, time
1...
20
...999 µL/min
0...
10
...999 s, INF
0...999 999 s,
OFF
>SET2
Parameters for EP2. Identical as SET1.
>titration parameters
titr.direction:
pause 1
start V:
start V
factor
dos.rate
pause 2
extr.time
dos.element
meas.input:
temperature
time interval
General titration parameters.
+:Titration to higher pH, voltage, or
current. Auto: Direction is set
automatically.
Waiting time
before
start volume.
Type of start volume: absolute or relative.
Volume for
absolute
start volume.
Factor for
calculation of
relative
start
volume: factor * smpl size.
Dosing rate for start volume.
Waiting time
after
start volume.
Extraction time.
Selection of the dosing unit.
Measuring input for pH and U.
Titration temperature.
Time interval for measured value
acquisition.
+, -,
auto
0
...999 999 s
abs., rel.,
OFF
0
...999.99 mL
0
...±999 999
0.01...150 mL/min,
max.
0
...999 999 s
0
...999 999 s
internal D0
,
external D1/D2
1
, 2, diff.
-170.0...
25.0
...500.0 °C
1...
2
...999 999 s
>stop conditions
stop V:
stop V
factor
filling rate
Type of stop volume: absolute or relative.
Volume for
absolute
stop volume.
Factor for
calculation of
relative
stop
volume: factor * smpl size.
Filling rate after titration.
abs.
, rel., OFF
0...
99.99
...9999.99 mL
0...±
999 999
0.01...150 mL/min,
max.
>statistics
Statistics calculation see at DET, page 14.
>preselections
conditioning:
display drift:
drift corr:
drift value
req.ident:
req.smpl size:
limit smpl size:
activate pulse:
Automatic conditioning of titration vessel
Display of drift during conditioning.
Type of drift correction.
Value for manual drift correction.
Request of identifications after start.
Request of sample size after start.
Limit control of sample size.
Pulse output on I/O line L6.
ON,
OFF
ON
, OFF
auto, man.,
OFF
0.0
...99.9 µL/min
id1, id1 & 2, all,
OFF
value, unit, all,
OFF
ON,
OFF
first, all, cond.,
OFF