3.4
Configuration, key <CONFIG>
799 GPT Titrino: Other functions
28
3.4 Configuration, key <CONFIG>
Display
Meaning Input
range
>monitoring
validation:
time interval
time counter
calibration:
meas.input:
time counter
service:
next service
system test report:
Monitoring functions
Validation of Titrino.
Time interval for the validation.
Time elapsed since the validation.
PH calibration
Measuring input.
Time interval for the calibration.
Service of instrument.
Date for the next service.
Print-out of the system test report after
switching on the Titrino.
ON,
OFF
1...
365
...9999 d
0
...9999 d
ON,
OFF
1
, 2, diff.
0
...9999 d
ON,
OFF
YYYY-MM-DD
ON,
OFF
>peripheral units
send to COM1:
send to COM2:
man.reports to COM:
balance:
stirrer
control:
remote
box:
keyboard:
barcode:
Settings of peripheral units.
Selection of printer at COM1.
as for COM1.
Output of manually triggered reports.
Selection of balance.
Stirrer control in titration sequences.
Connection of a remote box.
Type of the PC keyboard connected to
the remote box.
Target for datas eneterd by a barcode
reader.
Epson, Seiko, Citizen, HP,
IBM
1
, 2, 1&2
Sartorius
, Mettler, Mettler AT,
AND, Precisa
ON,
OFF
ON,
OFF
US, deutsch, francais, espanol,
schweiz.
input
, method, id1, id2, id3, smpl
size
>auxiliaries
dialog:
date
time
run
number
auto
start
start
delay
result
display:
dev.label
program
General settings.
Selection of dialog language.
Current run number for result output.
Automatic starts of titrations.
Waiting time before start of titration.
Result display at the end of
determination.
Device label.
Program version.
English
, deutsch, français,
español, italiano, portugese,
svenska
YYYY-MM-DD
HH:MM
0
...9999
1...9999,
OFF
0
...999 999 s
bold
, standard
up to 8 characters
read only
>RS232 settings COM1
baud
rate:
data
bit:
stop
bit:
parity:
handshake:
RS232 settings for COM1.
Baud rate.
Data bit.
Stop bit.
Parity.
Handshake.
300,600,1200,2400, 4800,
9600
,
19200, 38400, 57600, 115200
7,
8
1
, 2
none
, odd, even
HWs
, Swchar, Swline, none
>RS232 settings COM2
as for COM1.
>common variables
Values of common variables C30...C39.