![Metrohm 799 GPT Titrino Short Instructions For Use Download Page 25](http://html1.mh-extra.com/html/metrohm/799-gpt-titrino/799-gpt-titrino_short-instructions-for-use_1779570025.webp)
2.10
Calculations
799 GPT Titrino: Titration, dosing and measuring modes
23
2.10 Calculations
The results are calculated with the formulas given in the
method. For the calculation, the raw values, determined in
the method (EP's and C variables), are available.
The results and the raw values can be assigned for further
calculations, e.g. for statistics or as common variables.
Formula input and assignments, key <DEF>
Display Meaning
Input range
>formula
RS?
RS1=EP1*C01/C00
RS1 text
RS1 decimal places
RS1 unit:
RS1 limit control:
RS1 low lim.
RS1 up.lim.
RS1 L13 output:
Input of formula.
Enter result number.
Enter formula by means of 3
rd
functions of
keyboard.
Text for result output.
Number of decimal places for result
output.
Select result unit.
Limit control of the result.
Lower limit for RS1.
Upper limit for RS1
Output on line L13 if the result is out of
limits.
Enter values of calculation variables with
<C-FMLA>.
1...9
RS1
or up to 8 ASCII characters
0...
2
...5
%
, ppm, g/L, mg/mL, mg/pc,
mol/L, mmol/L, g, mg, mL, s,
mL/min, no unit or up to 6 ASCII
characters
ON,
OFF
0.0
...999 999
0.0...
999 999
OFF
, active, pulse
>silo calculations
C24=
C25=
match id:
Allocations for silo calculation, see page
96 of the Instructions for Use.
Enter values to be stored in the silo
memory.
Indication which id's must coincide for the
statistical evaluation in the silo memory.
RSX, EPX, CXX
id1, id1 & 2, all,
OFF
>common variables
C30=
Allocations of common variables
C30...C39.
Assign RSX, EPX, CXX, or MNX.
RSX, MNX, EPX, CXX
>report
report COM1:
report COM2:
Selection of report blocks for data output
at COM1.
Depends on the selected mode. If you
wish several reports, use ”;” as separator.
As for COM1.
full, short, mplist, curve, derive,
comb, meas crv, temp crv, scalc
full, scalc srt, calc, param, calib,
ff
>mean
MN1=
Allocations for statistics.
Assign RSX, EPX, or CXX.
RSX, EPX, CXX
>temporary variables
C70=
Allocations of temporary variables for
calculations under TIP.
RSX, EPX, CXX