1.4 Working with several dosing units
799 GPT Titrino: Getting started
11
Manual dosing
•
to prepare an Exchange Unit (or Dosing Unit, resp.)
•
for quick dosing
If you press the key <DOS>, the dosing unit from the
method is working. This dosing unit appears in the dis-
play.
If you wish to dose with another dosing unit, use the key
<PREP> to select the dosing unit and press <DOS> to
dose.
<PREP>
<PREP>
<DOS>, <STOP/FILL>
<QUIT>
Press <PREP>. In the display appears "inter-
nal D0 prep", i.e. the dosing unit of the Titrino has been
selected.
Press <PREP> again. In the display appears "exter-
nal D1: prep".
Press the key <DOS> on the Titrino: The external dos-
ing unit D1 doses. If you press the key <STOP/FILL> on
the Titrino, it will fill again.
With <START>, the function "prep" is started. This pre-
pares the dosing units. (The parameters for "prep" are
under the key <CONFIG>, in the group ">prep.dosing
elements").
Attention:
With the function "prep", liquid will be ex-
pelled!
If the dosing unit D1 is a Dosino, you can empty the dos-
ing unit completely (e.g. for cleaning). Choose the func-
tion "empty" with the key <
←
> or <
→
> and start with
<START>.
Attention:
With the function "empty", liquid will be ex-
pelled!
Quit the selection of dosing units with <QUIT>.
internal D0 prep