6.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
50
■■■■■■■■
942 Extension Module Vario SeS/PP (2.942.0500)
Problem
Cause
Remedy
Baseline increasing
strongly.
Insufficient flow of regener-
ation solution or rinsing
solution.
See problem "MSM – Insufficient flow of
regeneration solution or rinsing solution."
See problem "SPM – Insufficient flow of regen-
eration solution or rinsing solution."
MSM – Reduced capacity.
Regenerate the MSM
SPM – Reduced capacity.
Regenerate the SPM
Precision problems -
the measured values
are highly scattered.
MCS – The vacuum is too
low.
■
Check the connectors. If they are OK:
■
Contact Metrohm Service.
SPM – Insufficient
amount of regenera-
tion solution or rins-
ing solution is being
pumped.
There is a leak in the sys-
tem.
Check all connections.
Peristaltic pump – Contact
pressure too weak.
Correctly set the contact pressure
the contact pressure correctly", page 20)
.
Peristaltic pump – Filter
blocked.
Replace the filter
SPM – Backpressure too
high.
Clean the SPM
(see chapter 5.3.3.4, page 43)
or replace parts
Peristaltic pump – Pump
tubing defective.
Replace the pump tubing.