5.3 Metrohm Suppressor Module (MSM)
■■■■■■■■■■■■■■■■■■■■■■
46
■■■■■■■■
942 Extension Module Vario SeS/PP (2.942.0500)
Replacing parts of the Metrohm Suppressor Module (MSM)
1 Disconnecting the Metrohm Suppressor Module (MSM) from
the IC system
■
Switch off the instrument.
■
Disconnect all capillaries of the MSM from the IC system.
2 Dismantling the Metrohm Suppressor Module (MSM)
■
Unscrew the union nut
■
Pull the connecting piece
out of the housing together with
the rotor
If the rotor gets stuck in the housing, you can push it out as fol-
lows:
Put a pointed object into the slot in the housing and use it to
push out the rotor.
■
Loosen the connecting piece from the rotor with a turning move-
ment.
3 Cleaning the new rotor
■
Clean the sealing surface of the new rotor
with ethanol
using a lint-free cloth.
4 Inserting the new rotor
CAUTION
The rotor may be destroyed during start-up if it is not inserted cor-
rectly.
NOTICE
(see "Inserting small rotors", page 23)
is
required in order to insert a small rotor into the suppressor drive.