5 Operation and maintenance
942 Extension Module Vario SeS/PP (2.942.0500)
■■■■■■■■
45
NOTICE
The rotor's sealing surface is located approx. 4 mm deep inside the
suppressor drive if the rotor is inserted correctly.
If this is not the case, then the rotor must be moved into the cor-
rect position using careful turning. If the rotor cannot be turned or
removed, it can be moved into the correct position from below by
means of a pointed object (e.g. a screwdriver).
6 Cleaning the connecting piece
■
Clean the sealing surface of the connecting piece
with
ethanol using a lint-free cloth.
7 Inserting the connecting piece
See also Chapter 3.5.1, page 21
■
into the housing in such a way
that connector
1
is on top and the three pins of the connecting
piece fit into the corresponding recesses on the housing.
■
Reattach the union nut
and tighten by hand (do not use a
tool).
8 Connecting and conditioning the Metrohm Suppressor Mod-
ule (MSM)
■
Reconnect the MSM to the IC system.
■
Before switching the MSM over for the first time, rinse each of the
three suppressor units with solution for five minutes.
5.3.3.5
Replacing parts of the Metrohm Suppressor
Module (MSM)
Parts of the Metrohm Suppressor Module (MSM) may need to be replaced
in the following cases:
■
Irremediable loss of suppressor capacity (reduced phosphate sensitivity
and/or significant rise in the baseline).
■
Irremediable blockage of the MSM (solutions can no longer be pum-
ped through the MSM).
Both the rotor and the connecting piece can be replaced.