■■■■■■■■■■■■■■■■■■■■■■
3 Installation
2.850.9110 IC Amperometric Detector
■■■■■■■■
13
In the
Workplace
program part:
■
Load the method.
■
In the
Watch window
, display the
Current
channel with at least
three decimal places.
5 Carrying out the test
In the
Manual
program part:
■
Set all components of the IC instrument to inactive, particularly
the high-pressure pump(s).
■
On the tab of the detector, switch on the dummy cell with
[Apply]
.
After no more than one minute, the detector signal should level
off at 2.667 nA ± 7%. Noise should not exceed 0.005 nA.
■
Switch the dummy cell off with
[Cell Off]
.
With the dummy cell switched off and the detector hardware still
running, the signal should drop below an absolute value of 1 nA,
and noise should be limited to the third decimal place.
Exactly even signals may indicate that new detector data is not
correctly transmitted.
6 Removing the dummy cell
■
Pull out the electrode connection cables from the connectors
AE
,
WE
and
RE
of the dummy cell.
■
Remove the dummy cell from the tray.
The dummy cell incorporates a resistor (300 M
Ω
) and a capacitor (100 nF)
connected in parallel. If, in DC mode, a potential of 0.8 V is applied, then
a current of 2.667 nA (± 7%) is measured in the dummy cell. The capaci-
tor simulates a well-working measuring cell.
3.3.2
Testing the leak sensor
The leak sensor should not respond during the start-up. If the leak sensor
nevertheless does respond during the start-up, you will find information
for eliminating the problem in Chapter
To check whether the leak sensor is functioning, proceed as follows:
Testing the leak sensor
1
Hold a cloth moistened with eluent or tap water on the two contacts
of the leak sensor
.
The leak sensor of the detector responds.
If the leak sensor does not respond, please request Metrohm Service.
Summary of Contents for 2.850.9110
Page 1: ...IC Professional Detector IC Amperometric Detector Manual 8 850 8062EN 2020 09 25 ...
Page 2: ......
Page 8: ......