3.3 Putting the amperometric detector into operation
■■■■■■■■■■■■■■■■■■■■■■
12
■■■■■■■■
2.850.9110 IC Amperometric Detector
■
The three electrode connection cables (6.2165.000)
1 Connecting the electrode connection cables to the dummy
cell
■
Plug the angled plug of the working electrode connection cable
(labeled
WE
) into the
WE
socket.
■
Plug the angled plug of the reference electrode connection cable
(labeled
RE
) into the
RE
socket.
■
Plug the angled plug of the auxiliary electrode connection cable
(labeled
AE
) into the
AE
socket.
2 Connecting the electrode connection cables to the detector
(unless they are already connected)
■
Plug the straight plug of the working electrode connection cable
(red sleeve) into the
WE
socket of the detector.
■
Plug the straight plug of the reference electrode connection cable
(black sleeve) into the
RE
socket of the detector.
■
Plug the straight plug of the auxiliary electrode connection cable
(blue sleeve) into the
AE
socket of the detector.
3 Inserting the dummy cell
■
Place the dummy cell into the tray of the detector.
■
Attach the front cover.
NOTICE
The metal parts of the cable plugs must not touch the front cover.
4 Adjusting settings in MagIC Net
In the
Method
program part, create a new method for the instru-
ment test with the dummy cell.
■
Add the detector and the instrument and select them.
■
Select the
DC
mode.
■
Set the following parameters for the DC mode:
–
DC potential
:
0.8 V
–
Range
:
Auto
–
Damping
:
off
■
Add an analysis for the detector channel
Current
.
■
Add the entry
Current
▶
Start data acquisition
in the Time
program subwindow.
■
Save the method.