3.5 Attaching the front cover
■■■■■■■■■■■■■■■■■■■■■■
20
■■■■■■■■
2.850.9110 IC Amperometric Detector
2
Plug the straight plug of the reference electrode cable (black sleeve)
into the
RE
socket of the detector.
3
Plug the straight plug of the auxiliary electrode cable (blue sleeve)
into the
AE
socket of the detector.
Connecting the electrode cables to the measuring cell
Prerequisites:
■
The working electrode and the reference electrode are inserted into the
measuring cell.
1
Plug the angled plug of the working electrode cable (labeled
WE
)
into the working electrode socket.
2
Plug the angled plug of the reference electrode cable (labeled
RE
)
into the reference electrode socket.
3
Plug the angled plug of the auxiliary electrode cable (labeled
AE
) into
the socket (labeled
AE
).
3.5
Attaching the front cover
In order to obtain good measuring results, we recommend that the front
cover be put back in place.
When you are attaching the front cover, observe the following:
■
Do not pinch any capillaries!
Guide the capillaries through the capillary feed-throughs
.
■
Do not pinch any cables!