3.2 Inserting the amperometric detector
■■■■■■■■■■■■■■■■■■■■■■
10
■■■■■■■■
2.850.9110 IC Amperometric Detector
Placing the detector in the instrument
1 Remove the knurled screws
■
Unscrew the knurled screws on the rear panel.
■
Optional with 850 instruments
: Remove the handle if it is still fas-
tened to the instrument.
2 Remove the rear panel
■
Remove the rear panel.
3 Position the detector
■
Put the detector through the opening and position it on the sup-
port surface intended for this purpose and slide it as far as the
limit stop to the front.
4 Replace the rear panel
■
Insert the detector cable in a cable feed-through on the rear
panel.
■
Replace the rear panel.
The detector cable must hang freely from the cable feed-through.
5 Tighten the knurled screws
■
Optional with 850 instruments
: Reattach the handle higher up
and use it as a holder for MPaks.
■
Tighten the knurled screws.
Connecting the detector connection cable
CAUTION
The IC instrument
must
be switched off when the detector is being
connected.
Before connecting the detector, make sure that the IC instrument is
switched off.