3 Installation
2.850.9110 IC Amperometric Detector
■■■■■■■■
17
2 Inserting the measuring cell
Insert the chip of the measuring cell into the cell holder so that you
can hear it locking in.
NOTICE
Do not move the measuring cell for at least 5 seconds after having
inserted it.
During this time, data is read from the chip of the measuring cell
and written into the database. This process must not be interrup-
ted, because otherwise the data may be transferred incorrectly or
incompletely.
3 Testing at low flow
■
In the program part
Manual
of MagIC Net, set the flow rate of
the high-pressure pump to 0.2 mL/min and start the high-pressure
pump.
■
Watch the detector outlet capillary: Liquid must emerge from the
end of the detector outlet capillary.
If no liquid emerges from the end of the detector outlet capillary:
– Detach the capillary from the
Out
connection on the mea-
suring cell and check whether the end has been pinched by
the pressure screw.
– Shorten the capillary and fasten once again to the
Out
con-
nector on the measuring cell.