MI 3295 Step Contact Voltage Measuring System
MI 3295S Station operation
20
6 MI 3295S Station operation
6.1 Organization of displays
Figure 6.1: Typical display in earth
resistance function
Function name
Result & Sub-result field
Message field
Figure 6.2: Typical display in
specific earth resistance function
Function name
2.4m
Test parameter field
Result & Sub-result field
Message field
Figure 6.3: Typical display in
current function
Function name
Result field
Message field
Remained place in current logger
6.2
Warnings and messages
Before and during the measurement the instrument performs different tests to ensure
safety and to prevent from damaging it. These safety tests include checking for any
external voltage or improper loading of test terminals. If a problem is detected, an
appropriate warning message will be displayed. Warnings and protective measures are
described in this chapter.
Voltage between test terminals C1/H and C2/E is higher than
allowed (>10 V). Disconnect the test leads and check why an
external voltage was detected!
Output was overloaded or test current abruptly fall down.
Overload can be caused by high spurious earth current. In this
case it is recommended to decrease the output power. See
chapter
Output current is too low. A too low current can be caused
because of bad connection or high probe resistance.