MI 3295 Step Contact Voltage Measuring System
Maintenance
44
Fuse replacement
Figure 9.2: Fuse replacement procedure
There are two fuses on the front cover of the MI 3295S Station:
Fuse: 2 x T 6.3 A / 250 V, (5 mm
20 mm)
Fuses prevents from danger in case of a fault inside the instrument.
Warnings:
Disconnect all measuring accessories, switch off the instrument and
disconnect mains cord before opening the cover of fuse holder, hazardous
voltage inside!
Replace blown fuse with original type only, otherwise the instrument may be
damaged and/or operator’s safety impaired!
Position of a fuse can be seen in
9.2 Cleaning
No special maintenance is required for the housing. To clean the surface of both
instruments Meter (MI 3295M) and Station (MI 3295S) use a soft cloth slightly moistened
with soapy water or alcohol. Then leave the instrument to dry totally before use.
Warnings:
Do not use liquids based on petrol or hydrocarbons!
Do not spill cleaning liquid over the instrument!