MD 9260 Clamp Meter
Sicherheit
15
Electrical Specifications
Accuracy is
±
(% reading number of digits) or otherwise specified, at 23
°
C
±
5
°
C & less
than 75% relative humidity. True RMS voltage accuracies are specified from 5 % to 100 % of
range or otherwise specified. Maximum Crest Factor < 1.4 : 1 at full scale & < 2.8 : 1 at half scale,
and with frequency components within the specified frequency bandwidth for non-sinusoidal
waveforms.
DC Voltage
RANGE
Accuracy
6.000V, 60.00V,
600.0V & 1000V
0.5%+5d
1000 V–1500V
0.5%+5d
Input Impedance: 10M
Ω
, 50 pF nominal
AC Voltage
RANGE
Accuracy
50Hz ~ 400Hz
6.000V, 60.00V,
600.0V & 1000V
1.2% + 5d
Input Impedance: 10M
Ω
, 50 pF nominal
AC+DC Voltage
RANGE
Accuracy
DC, 50Hz ~ 400Hz
6.000V, 60.00V,
600.0V & 1000V
1.4% + 7d
Input Impedance: 10M
Ω
, 50 pF nominal
VFD_ACV (with Low Pass Filter )
RANGE
Accuracy
1)
10Hz ~ 20Hz
6.000V, 60.00V,
600.0V & 1000V
4%+80d
20Hz ~ 200Hz
6.000V, 60.00V,
600.0V & 1000V
2%+60d
200Hz - 400Hz
2)
6.000V, 60.00V,
600.0V & 1000V
7%+80d
1)
Not specified for fundamental frequency > 400Hz
2)
Accuracy linearly decreases from 2% + 50d @ 200Hz to 7% + 80d @ 400Hz
CREST-MAX Capture Mode
Accuracy: Specified accuracy plus 250 digits for changes > 5ms in duration
Ohm & AutoCheck
TM
_Ohm
1)