MD 9260 Clamp Meter
Sicherheit
14
5.
Specifications
General
Display:
3-5/6 digits 6000 counts. & 3-1/2 digits 1,999 counts for Hz
Polarity:
Automatic
Update Rate:
5 per second nominal
Operating Temperature:
0
°
C to 40
°
C
Relative Humidity:
Maximum relative humidity 80% for temperature up to 31
°
C
decreasing linearly to 50% relative humidity at 40
°
C
Pollution degree:
2
Storage Temperature:
-20
°
C to 60
°
C, < 80% R.H. (with battery removed)
Altitude:
Operating below 2000m
Temperature Coefficient: nominal 0.15 x (specified accuracy) /
°
C @(0
°
C–18
°
C or 28
°
C–40
°
C),
or otherwise specified
Sensing:
True RMS
Safety:
Meets IEC/EN/BSEN/CSA_C22.2_No./UL standards of 61010-1 Ed.
3.1 and 61010-2-032 Ed. 4.0 to Measurement Categories CAT IV
1000V ac&dc and CAT III 1500Vdc.
Transient Protection:
12kV (1.2/50
µ
s surge)
Overload Protections
Clamp-on jaws:
2000A rms continuous “+“ & COM
Voltage on terminals:
1650Vdc / 1100Vac rms
Other functions:
1500Vdc / 1000Vac rms
E.M.C.:
Meets EN61326-1:2006 (EN55022, EN61000-3-2, EN61000-3-3,
EN61000-4-2, EN61000-4-3, EN61000-4-4, , EN61000-4-5,
EN61000-4-6, EN61000-4-8, EN61000-4-11)
In an RF field of 3V/m: Capacitance function is not specified
Other function ranges: Total Accuracy = Specified Ac 200 digits
Performance above 3V/m is not specified
Power Supply:
1.5V AA Size (IEC LR6) battery X 2
Power Consumption:
Typical 14mA for Current functions, and 5.2mA for others
Low Battery:
Below approx. 2.4V
APO Timing:
Idle for 34 minutes
APO Consumption:
10
µ
A typical
Dimensions:
L264mm X W97mm X H43mm
Weight:
608 g
Jaw opening/Conductor
diameter:
55mm max
Accessories:
Test leads (pair), user's manual, banana plug K-type
thermocouple x 1
Optional purchase
accessories:
USB interface kit AMD 9250; AMD9024 banana plug to type-K
socket plug adaptor
Special Features
AutoCheck
TM
V&
Ω
; VFD-V & VFD-Hz; Backlighted LCD; 5ms
CREST-MAX Capture mode (Peak Hold); Auto-ranging Relative-
Zero mode; Display Hold; EF-Detection (NCV); Optional Interface
capabilities with PC computers