Connection for Uoc / Isc measurement
Figure 5.35: Connection with universal PV test lead
Figure 5.36: Connection with PV safety probe
Uoc / Isc measurement procedure
Select
Uoc / Isc
sub-function using the function selector keys and
/
keys.
Connect
universal PV test lead / PV safety probe and sensors (optional) to the
Connect
accessories to the item to be tested (see
).
Check value and polarity of input voltage.
Press the
TEST
key to perform the measurement.
Store
the result by pressing the
MEM
key (optional).
Figure 5.37: Example of Uoc / Isc measurement results
Displayed results for Uoc / Isc measurement:
MEAS column
Uoc .......... measured open voltage of the panel