MI 3109 EurotestPV Lite
Data handling
59
6.5.4 Renaming installation structure elements (upload from PC)
Default installation structure elements are “Object”, “Inverter”, “String” and ‘’Panel’’.
In the PCSW package Eurolink-PRO default names can be changed with customized
names that corresponds the installation under test. Refer to PCSW Eurolink-PRO
HELP for information how to upload customized installation names to the instrument.
Figure 6.12: Example of menu with customized PV installation structure names
6.5.5 Renaming installation structure elements with serial barcode
reader or RFID reader
Default installation structure eleme
nts are “Object”, “Inverter”, “String” and “Panel”.
When the instrument is in the Save results menu location ID can be scanned from a
barcode label with the barcode reader or can be read from a RFID tag with the RFID
reader.
Figure 6.13: Connection of the barcode reader and RFID reader
How to change the name of memory location
Connect the barcode reader or RFID reader to the instrument.
Make sure that RS232 is selected in Communication menu.
In Save menu select memory location to be renamed.
A new location name (scanned from a barcode label or a RFID tag) will be
accepted by the instrument. A successful receive of a barcode or RFID tag is
confirmed by two short confirmation beeps.
Note
:
Use only barcode readers and RFID readers delivered by Metrel or authorized
distributor.