OmegaPATPlus / BetaPATPlus Measurements
34
6.5 Touch
leakage
current
This test determines the current that would flow if a person touches the appliance.
Measurement is based on a human body model with resistance of 2 k
Ω
. Both AC and
DC components of touch leakage current are detected.
The instrument measures the leakage current flowing through the S probe into earth.
The appliance under test can be powered from the mains test socket or directly from the
installation (fixed installed equipment).
Measurement of touch leakage current
Measurement of touch leakage current on a fixed installed appliance
6.5.1 How to perform single touch leakage current measurement
From the
Main menu
, select
Single test
and then select
Touch leakage
by using
¿
and
À
cursor keys and press ENTER key to confirm.
Touch leakage menu
is
displayed.
Touch leakage menu