OmegaPATPlus / BetaPATPlus Instrument description
9
13............Alpha-numeric keyboard.
14............STOP key.
15............START key.
16............ESCAPE key.
17............Current clamp adapter input sockets.
Warning! Do not connect any voltage source on this input. It is intended
only for connection of current clamp with current output. Maximum
input current is 30 mA!
18............Function keys.
19............USB connector.
20............PC / printer connector.
21............Barcode reader connector.
22............Label printer connector on back side (support for PrintekMobile MtP300 /
MtP400 serial printer).
Safety pre-tests
Before performing a measurement, the instrument performs a series of pre-tests to
ensure safety and to prevent any damage. These safety pre-tests check for:
Ì
any external voltage against earth on test socket,
Ì
excessively high leakage current,
Ì
excessively high touch leakage current,
Ì
short circuit or too low resistance between L and N on appliance,
Ì
correctly applied mains voltage to the test socket.
If pre-tests fail, an appropriate warning message will be displayed.
The warnings and measures that have to be taken are described in chapter
2.2
Meaning of symbols and messages on the instrument display
.
2.2 Meaning of symbols and messages on the instrument
display
Mains voltage is not
correct or PE is not
connected.
Check mains voltage
and PE connection!
Improper supply voltage warning. Possible causes:
Ì
no earth connection or other wiring problem on
supply socket.
Ì
incorrect mains voltage.
Determine and eliminate the problem before
proceeding!