OmegaPATPlus / BetaPATPlus
Autotest sequences
55
7.4 Performing autotest sequences
7.4.1 Visual inspection
During a visual inspection of an appliance damage to the following items should be
identified:
Ì
appliance
plug,
Ì
appliance
mains
fuse,
Ì
appliance
mains
lead.
Visual test menu
In case that any damage on the appliance is identified, press F3 key (FAIL) in order to
identify the visual test as failed.
7.4.2 Earth bond resistance measurement
Note:
Ì
Consider the displayed warnings before starting measurement!
If the earth bond resistance measurement passes, the instrument automatically
proceeds with the next test in the sequence (in single measurement mode only).
If the earth bond test fails or if the earth bond is set to continuous mode, the autotest
sequence stops when the earth bond resistance measurement is performed and a
prompt will appear on the screen.
The following actions can be taken to continue or complete the autotest sequence:
START...................... Starts the earth bond resistance measurement.
Proceed with the next earth bond resistance measurement (in
continuous measurement mode only).
Proceed to the next autotest sequence measurement (in single
measurement mode only).
F2 function key...........View the earth bond test help screens.
F3 function key ......... Repeat the earth bond resistance measurement.
ENTER key............... Proceed to the next autotest sequence measurement (in
continuous measurement mode only).
F4 function key ......... Skip earth bond resistance measurement.
F5 function key ......... End the autotest sequence.