4. The sample time is fixed at 60 seconds.
5. Status bits combinations are possible. See the Status Bits table.
7.4.2. Mass Format
Mass Data Report
2013-01-22 11:59:52
Serial Number, N12345
Time,PM1(µg/m3),PM2.5(µg/m3),PM4(µg/m3),PM7(µg/m3),PM10(µg/m3),TSP(µg/m3),AT(C),RH(%),Location,Seconds,Status
2013-01-22 11:23:27,001.1,003.8,005.2,005.9,006.2,008.2,+020,026,001,060,000
CSV Fields
Field
Parameter
Example Value
Notes
1
Date and Time
2013-01-22 11:23:27
2
PM1 (µg/m3)
001.1
3
PM2.5 (µg/m3)
003.8
4
PM4 (µg/m3)
005.2
5
PM7 (µg/m3)
005.9
5
PM10 (µg/m3)
006.2
6
TSP (µg/m3)
008.2
7
AT (C, F)
+021
Note 2 & 3
8
RH (%)
40
Note 3
9
Location
001
10
Seconds
60
Note 4
11
Status
000
Note 5
Status Bits
Bit
Value
Condition
0
OK (no alarms or errors)
0
1
Not used
1
2
Not used
2
4
Not used
3
8
Not used
4
16
Low battery
5
32
Sensor error
Notes (for table above)
1. CSV header included for multiple record transfers like All Data (2)
or New Data (3).
2. Units determined by product setting.
3. Temperature and RH will be null (,,) if Temp/RH probe is not
attached
4. The sample time is fixed at 60 seconds.
5. Status bits combinations are possible. See the Status Bits table.
For example, 48 (00110000B) = Low battery and Sensor error.