AEROCET 531S Manual Rev H
Page 15
5.1.
Edit Menu Items
To change settings, press
MENU
to display the main menu,
press ▲ or ▼ to navigate to
the desired item and press
ENTER
to display the item view/edit screen.
To edit pick list items (e.g.
SAMPLE SETUP MODE: AUTO
or
MANUAL
), press ▲ or ▼ to
navigate to the item. Press
ENTER
to select the item. Press ▲ or ▼ to change the
setting. Press
ENTER
to save the setting or
ESC
to cancel and return to the main value.
To edit numeric values (e.g.
LOCATION
), press ▲ or ▼ to navigate to the item. Press
ENTER
to select the item. Press ▲ or ▼ to increment or decrement a value. Press ◄ or
► to select the next digit. Press
ENTER
to save the value or
ESC
to cancel and return to
the main value.
5.2.
Sample Setup Screen
Figure 11 shows the
SAMPLE SETUP
Screen. The 4 parameters are covered in the
following sections.
LOCATION: 001
MEASURE: COUNTS
MODE: MANUAL
HOLD: 000
Figure 11
–
SAMPLE SETUP
Screen
5.2.1. Location Number
The
LOCATION
number is used to assign a unique number to a location or area. This
important field is included in sample data records (display, printer and csv output).
5.2.2. Measure
The
MEASURE
is either
MASS
or
COUNTS
concentration.
Selection
Description
MASS
When
MASS
is selected the unit will report PM levels
PM1, PM2.5, PM4, PM7, PM10 and TSP in µg/m
3
.
COUNTS
When
COUNTS
is selected the unit will report fixed
particles sizes
.
0.5 µm, 1.0 µm, 5.0 µm, and 10.0 µm when using
low sensitivity
0.3 µm, 0.5 µm, 1.0 µm, 5.0 µm, and 10.0 µm
when using high sensitivity.