Operating
the Software
Software
28
/
58
20.03.2015
8LE-D23_BA_E
5.3.5. Parameterization of M=Light LED for Pulsed Operation
Figure 15: Flash light window
Designation
Description
Trigger input
Selects the input signal for the trigger (Figure 17, page 33).
Pulse Count
Number of periods per trigger event. In the
Software (Mastermode)
operating mode an
unlimited number of pulses is emitted if the value is set to
0
.
Pulse period [µs]
Duration of a period (pulse + pause) in µs. If you set the value for
Pulse periode
to 0,
the
Auto-Follow
operating mode is switched on.
Pulse width [µs]
Duration of a pulse in µs (switch-on duration).
Pulse delay [µs]
Time delay by which the pulse is to be retarded in comparison with the trigger input
time. If for
Pulse delay
the parameterized value is 0, the delay between trigger event
and light flash is approx. 10µs.
Power [%]
Intensity with which M=Light LED is to be switched on as long as the pulse is active.
In pulsed operation, the intensity may be increased up to 255%.
Send
Transmit set parameters to M=Light LED.
Off
Switches all connected M=Light LEDs to power = 0%
Trigger
Theoretically possible pulse sequence at the trigger input for selected parameters.