and
SQA-V
Service Manual Dec_2019
40
“Empty I-button inserted”
Press ESC and Retry with a
new I-Button
XX denotes quantity of tests
remained
“I-button warning”
This screen will be displayed
when less than 10 tests remain
in the SQA-V
Insert a new I-Button
“Please insert a new I-button”
This screen will be displayed
when test new patient is
activated, but no more tests
remain in the SQA-V
Insert a new I-Button
7.
Communication error
A communication error may
occur in one of the following
cases:
Defective wire connection: The
cable that connects the optical
board to the main board is not
well connected or was
damaged
Hardware problem
EMPTY I-BUTTON INSERTED!
INSERT A NEW I-BUTTON
# TESTS REMAINING: XX
PRESS ESC TO RETRY
I-BUTTON WARNING:
LESS THAN 10 TESTS REMAINING
PLEASE INSERT A NEW I-BUTTON
NO MORE TESTS ARE AVAILABLE
COMMUNICATION ERROR
CALL FOR TECHNICAL SUPPORT