MES SQA-V Service Manual Download Page 12

     and 

SQA-V 

Service Manual Dec_2019     

 

 

 

12 

Electronic 

SECTION V: Electronic Self-Test and Auto-Calibration 

Self-Test 

and Auto-

Calibration  

A series of tests are automatically run on the SQA-V to check the calibration and 
internal operating system. Tests are run when the SQA-V is first turned-on and 
also prior to each semen analysis test.  

Start-up: 

 

Stabilization and auto calibration:

 Checks system stability and reference 

ranges. The system sensors are analyzed for several minutes to ensure that the 
values are within a very narrow acceptable range. Once the system is stable for 30 
seconds it will pass stabilization and auto calibration. The system will fail if it is not 
stable for at least 30 seconds and a warning message will be displayed.  

 

System noise:

 The level of electronic noise in the system is measured to ensure 

that noise thresholds are accurately defined to ensure effective measurement of 
electronic signals. The system will use this measurement prior to running a test 
and will automatically adjust the noise level thresholds to ensure accurate 
readings.  

 

Self-test:

 The system produces electronic signals that simulate motility and 

concentration in order to check the performance of the system and verify that the 
calibration settings are consistent with the factory specifications. The  
SQA-V will report failures (see section on error and warning messages) and 
"freeze" the system if the system is not within the established ranges.  

Prior to testing a sample: 

 

Auto calibration verification

: Reference values are read again. The 

concentration and motility parameters are measured (without a testing capillary).  

 

System noise:

 Measures the electronic noise level of the system to ensure that 

noise thresholds are accurately defined in the system to ensure effective 
measurement of electronic signals. The system will use this measurement prior to 
running a test and will automatically adjust the noise level thresholds to insure 
accurate readings.  

 

Electronic spikes:

 Checks for any measurement points that are out of range 

electronically. More than three such points will fault the system and a warning 
message will be displayed. 

 
The following procedure details how the user can document the system 
parameters to prepare for technical support if the system fails (see section of this 
user guide on Error Messages and Warning Messages).  

 
How to run a copy of the system parameters version 2.48/2.49 GOLD:  

 

From the 

MAIN MENU

 select:  

SERVICE>PRINT SQA-V DEFAULT 

SETTINGS>SELF TEST DATA 

 
All other versions: 

 

Remove any testing capillaries from the system. 

 

Turn the SQA-V on from rear panel and from the front keypad. 

 

When the MAIN MENU appears or when a FAILED SELF TEST message appears 
press the 

SERVICE

 key on the keypad to enter the SERVICE menu. 

 

Select 

USER

 

Press the 

PRINT

 key on the keypad to generate the required printout. 

 

Refer to the table below. Enter numbers in the "SQA-V Value" column that 
corresponds to #1 - #10 from the SQA-V system parameters printout. Compare 
the values. If the value from the SQA-V is within range mark the "Pass" column. 
If not, mark the "Fail" column. 

 

Summary of Contents for SQA-V

Page 1: ...and SQA V S e r v i c e M a n u a l V e r s i o n 2 4 8 2 4 9 2 6 0 Catalog 7501 Rev Dec 2019...

Page 2: ...roubleshooting Overview 13 SECTION VIII Technical Support and Part Replacement Instructions Opening the SQA V 13 Closing the SQA V 14 Printer Thermal 14 Printer Impact 15 Keypad 15 Power Supply PSU 16...

Page 3: ...and SQA V Service Manual Dec_2019 3 Communication Error 40 SECTION X Appendixes Cleaning Instructions 41 Loading I Button Tests 42 I Button troubleshooting 43 Power Supply Troubleshooting Flowchart 44...

Page 4: ...perm ejaculate FUNC SPERM M Motile Sperm Concentration MSC M ml Total Morphologically Normal Sperm ejaculate MORPH NORM SPERM M Progressively Motile Sperm Concentration PMSC M ml Postvasectomy Motile...

Page 5: ...mm section of the capillary by analyzing light modulations caused by sperm motion Step 4 This information is then digitized and routed to the microprocessor that applies algorithms to extract the req...

Page 6: ...cus knob Display s Operational LCD screen with white backlight resolution 240 x 128 pixels Video LCD screen 4 diagonal resolution 960 x 240 pixels Use monitor On Off key on the keypad to independently...

Page 7: ...photometer channel input signal Modulated 1 kHz analog up to 5V Quality Control Internal Electronic Self Test and Auto Calibration External Counts can be run on the SQA V automated and visualization s...

Page 8: ...ions Operate the SQA V away from devices that may cause electronic noise cell phones or other devices causing vibrations such as centrifuges Turn system OFF at the rear panel when not in use for exten...

Page 9: ...eld Use the ESC button to return to the previous screen or field Rear Panel Side Panel The side panel has an I Button port with clip in newer SQA V s starting with Serial Number 1333 Refer to the Appe...

Page 10: ...alization Compartment Visualization Compartment Cover Measurement Compartment Focus control knob Schematic of the SQA V 1 5 to 1 8 VDC 5 VDC 5 VDC 12 VDC 3 VDC 12 VDC 12 VDC 2 VDC AC 100 230 V 50 60 H...

Page 11: ...ion Slide Adaptor For use with a standard laboratory slide 76 x 25 6 mm and 22 x 22 mm cover slip in the visualization compartment of the SQA V to visually assess semen samples Cleaning Kit Daily clea...

Page 12: ...d ranges Prior to testing a sample Auto calibration verification Reference values are read again The concentration and motility parameters are measured without a testing capillary System noise Measure...

Page 13: ...V without voiding the warranty SECTION VII Troubleshooting Overview The following sections describe how to troubleshoot and repair potential problems that may occur with the SQA V Theses section will...

Page 14: ...heck the cable connected between the printer and the main board location J3 and verify that is not damaged or disconnected If the cable is O K but printer is not working please proceed to printer repl...

Page 15: ...ne Reconnect the cable to the printer and main board location J3 Attach the printer to the SQA V rear panel Close the SQA V according to the instructions from Closing the SQA V section above Keypad 5...

Page 16: ...the Appendix section for instructions on how to address power supply problems 6 Power Supply Unit PSU ISSUE The main switch is ON but the power indicator does not light up and the fan is not working C...

Page 17: ...ement of the Power Supply Unit ISSUE The power supply and or cables are not working PLEASE NOTE The instructions depend on the SERIAL NUMBER SN of the SQA V SN 1329 and below follow the instructions i...

Page 18: ...onnector location CN1 Output connector location CN2 4 Using a 1 5 Philips screwdriver open four screws holding the faulty PSU 5 Replace the faulty PSU 6 Secure the new PSU to the rear panel using the...

Page 19: ...r SQA V of SN 1330 and above use part V H 00575 00 Note When replacing the fan do not change the direction of airflow Install the new fan in the same direction and orientation as a replaced fan 1 Open...

Page 20: ...CD screen Using a Phillips screwdriver remove the two screws that secure the faulty inverter board Replace the faulty inverter board with a new one and secure it with the two screws Re connect the cab...

Page 21: ...h a new one and secure it with the four screws 6 Reconnect the data flat cable to new Operation monitor 7 Reconnect backlight power cable to Inverter board location J2 SECTION 2 SN 2229 and above Note...

Page 22: ...orientation If data flat cable appears damaged in any way it should be replaced Item P N KHD 908 000858 4 If replacing the cable will not resolve the issue replace the processor on the main board see...

Page 23: ...onitor above 7 If the problem persists with the new screen perform an MBOB replacement OR send the SQA back to the manufacturer MES for a repair RMA SECTION 3 Replacing the SQA V processor Instruction...

Page 24: ...ocessor according to the following directions Align the dot which is marked on the processor with the Arrow mark on the processor socket see the figure to the right 5 Slightly press the center of the...

Page 25: ...the SQA V and disconnect the optical assembly from the front panel four screws Disconnect the focus knob from the optical assembly by unscrewing three screws Remove the old focus knob Insert the new f...

Page 26: ...he light source can be seen from the bottom of the LED house If no light can be seen please check the Illumination settings of SQA V 2 Illumination setting can be adjusted using designated buttons on...

Page 27: ...st the video settings in V Sperm Go to VIDEO Real Time Video Settings Check to make sure the Brightness and Contrast setting are adjusted to achieve optimal image Check the external cables connecting...

Page 28: ...of this manual If the problem persists after replacing the cable contact MES support 5 The image on the SQA V is frozen or stuck This type of problem can only occur in old systems with SN below 1318 I...

Page 29: ...ng the two plastic clips push the internal video cable connector inside to release it 8 Gently lay the optical assembly on its rear section 9 Move the internal video cable aside and make sure nothing...

Page 30: ...e lid fig 2 Do not start this procedure unless you have this track Contact MES if you need a FALSE TRACK 5 Unscrew the 2 screws holding DC Motor and remove it fig 3 and 4 6 Disconnect the internal vid...

Page 31: ...ns for CCD cleaning but this time do not remove the dust sleeve 2 Using a Phillips screwdriver remove the four screws holding the camera PCB 3 Replace the defective CCD camera with a new one 4 Fasten...

Page 32: ...dvised Only an experienced and licensed electrical technician or electrician should perform this repair How to check the Surge Protector Turn off the power and unplug the power cord from the inlet on...

Page 33: ...located on the front face of the surge protector and verify that the surface inside the windows is green If at least one of the surfaces inside the windows is not green replace the surge protector Ho...

Page 34: ...pillary sensor is defective and must be replaced follow the instructions in next section If the SELF TEST keeps failing while the values of Conc 1 Conc 2 and Conc 3 are ZERO call technical support for...

Page 35: ...l Board to the LED Board were damaged and therefore require replacement Note The following instructions for replacing the Optical Board cables are applicable only for devices where aforementioned cabl...

Page 36: ...Part V M00571 00 Issue A Visualization System Upgrade is available for purchase and can be easily installed on existing systems The upgrade consists of 3 components 1 Improved Visualization LED Housin...

Page 37: ...e the Focus knob See section 13 in this manual for detailed instructions Stage 4 Close the SQA V Re connect the optical assembly to the front panel Make sure to screw back all 4 screws Close the back...

Page 38: ...onic noise cell phones etc Clean measurement compartment refer to Appendix Reboot the SQA V without a testing capillary in the chamber Turn the system OFF then back ON at the main switch on the rear p...

Page 39: ...ration Out of Range A message will appear indicating that the tests results for SPERM CONC and or MSC are beyond the dynamic range established by the manufacturer for testing This message will appear...

Page 40: ...is activated but no more tests remain in the SQA V Insert a new I Button 7 Communication error A communication error may occur in one of the following cases Defective wire connection The cable that c...

Page 41: ...and SQA V Service Manual Dec_2019 41 Appendix 1 SQA Cleaning Instructions SECTION X Appendixes...

Page 42: ...button port of the SQA The back of the card instructions side should be facing up so that the clip touches the I button through the round hole at the back of the card See Fig 1 The I button should be...

Page 43: ...k CONTINUE again if the I button it doesn t load tests and you receive a communication error 2 If you still receive a Communication Error check the following Is the I button properly located fully ins...

Page 44: ...solved Yes Yes Yes Yes End of troubleshooting Not Not Not Not Voltage is OK Check the input voltage of PSU see p 16 Not OK Not OK Not OK Not OK Replace the power inlet see p 19 OK OK OK OK Check the...

Reviews: