403321001_001 2016-08-31 

 

   

 

MetaSTAR RF Generator User Manual 

Catalog No. 3195 (English) (EN) 

CAUTION: Federal law (US) restricts this device to sale by or on the order of a physician. 

1. 

BRIEF DEVICE DESCRIPTION: 

The  MetaSTAR RF Generator is  a microprocessor  controlled low  power  radiofrequency (RF)  generator  used for 
percutaneous  delivery  of  bi-polar  radiofrequency  energy  intended  to  heat  the  target  tissue  of  the  SpineSTAR™ 
Ablation Instrument. 

2. 

INDICATIONS FOR USE: 

The  MetaSTAR  RF  Generator  is  indicated  for  palliative  treatment  in  spinal  procedures  by  ablation  of  metastatic 
malignant lesions in a vertebral body. 

3. 

CONTRAINDICATIONS: 

In  some  cases  the  system  may  interfere  with  normal  functions  of  some  types  of  implanted  pacemakers.  The 
MetaSTAR RF Generator is contraindicated for patients who have pacemakers or other electrical implants. Please 
refer to the SpineSTAR Ablation Instrument instructions for use for a more comprehensive list of contraindications 
regarding specific procedures.   

4. 

WARNINGS: 
Warning:

 

Do not ablate in painful osteoporotic vertebra without tumor. 

Warning:  

Do not use this device in patients without metastatic malignant lesions in a vertebral body. 

Warning: 

Do not use this device in patients with multiple myeloma, solitary plasma cytoma, or primary malignant 
lesions in the index vertebra. 

Warning:

 

The device is designed for tumor ablation; follow the IFU for the applicable device(s) used during the 
procedure. 

Warning:

 

Hazardous electrical output. This equipment is for use only by qualified medical personnel trained in the 
use of the MetaSTAR RF Generator. 

Warning:

 

Do not operate the unit in close proximity to volatile solvents such as methanol or alcohol, or in the 
presence of flammable anesthetics, as explosion may occur.   

Warning: 

Do not operate unit in moist environment, as a shock hazard may exist. If liquids have entered the unit, 
the MetaSTAR RF Generator must be returned to the manufacturer for testing prior to use. 

Warning: 

Interference produced by the operation of high-frequency equipment may adversely affect the operation 
of other electronic medical equipment such as monitors, imaging systems.

 

Warning: 

Failure of the equipment could result in an unintended increase of output power.  

Warning: 

Use of accessories and cables, other than those specified may affect system performance, resulting in 
increased emissions or decreased immunity of the system.

 

Warning:

 

Servicing other than replacing the fuse should be performed by the manufacturer. 

Warning:

 

Use only hospital grade power cord. 

Warning: 

In case of Generator failure switch off/unplug the power cord. 

Warning: 

Ablation  procedures  must  be  performed  under  fluoroscopic  image  guidance.  Do  not  perform  ablation 
without imaging as it may result in severe injury to patient. 

Warning: 

Do not remove data port cover or connect any equipment to the data port. 

Warning:

 

To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective 
earth. 

 

 

 

Summary of Contents for MetaSTAR 3195

Page 1: ...West Merit Parkway South Jordan Utah 84095 U S A 1 801 253 1600 U S A Customer Service 1 800 356 3748 Authorized Representative Merit Medical Ireland Ltd Parkmore Business Park West Galway Ireland 31...

Page 2: ...PPLIED 4 7 1 Accessories 4 7 2 Single Use Devices See table below for other devices required for Tumor Ablation 4 8 INSTRUCTIONS FOR USE 5 8 1 Set Up Error Bookmark not defined 8 2 MetaSTAR RF Generat...

Page 3: ...FU for the applicable device s used during the procedure Warning Hazardous electrical output This equipment is for use only by qualified medical personnel trained in the use of the MetaSTAR RF Generat...

Page 4: ...istatic sheeting is recommended Caution Cables to the surgical electrodes are recommended to be positioned such that contact with patient or other leads is avoided Caution Do not continue to deliver R...

Page 5: ...SpineSTAR Ablation Instrument Hand Switch Cable andAE Cable are available as backup prior to use 8 3 MetaSTAR RF Generator Nomenclature Symbols ON power Type BF Applied Part Volume Control OFF power W...

Page 6: ...LOW LED 4 RF ON BLUE LED 5 POWER ON GREEN LED 6 Soft Key Button 1 7 Soft Key Button 2 8 Soft Key Button 3 9 Soft Key Button 4 10 Hand Switch Cable Socket 11 AE Cable Socket 12 Fuse Drawer 13 ON OFF Ma...

Page 7: ...be connected to the AE Cable socket on the MetaSTAR RF Generator using the AE Cable inspect the AE Cable for damage If the SpineSTAR Ablation Instrument is connected properly Distal and Proximal tempe...

Page 8: ...TAR Ablation Instrument is properly placed 1 Both thermocouple temperatures are displayed in amber 30 C Note temperatures are displayed in blue when temperature is 30 C 2 Electrode is extended to the...

Page 9: ...instructions for use and maintenance instructions The sole obligation of MERIT under this warranty shall be limited to the replacement of the MetaSTAR RF Generator or equivalent models at no charge t...

Page 10: ...erit Medical Systems Inc 1600 West Merit Parkway South Jordan Utah 84095 U S A 1 801 253 1600 U S A Customer Service 1 800 356 3748 Authorized Representative Merit Medical Ireland Ltd Parkmore Busines...

Page 11: ...operation Enclosure IP 21 III Operating Conditions Temperature 59 F to 104 F 15 C to 40 C Relative Humidity 30 to 75 non condensing Atmospheric Pressure 878 to 1082cmH2O 86 to 106kPa Whenever the tem...

Page 12: ...Instrument are intended for use in the electromagnetic environment specified below The customer or the user of the System should assure that it is used in such an environment Immunity test IEC 60601...

Page 13: ...ntific and Medical ISM RF Equipment MDD 93 42 EEC Medical Device Directive Council Directive Concerning Medical Devices APPENDIX E COMPATIBLE POWER CORDS Continental Europe 2 0m 10A Inter Power PN 862...

Page 14: ...in a particular installation If this equipment does cause harmful interference to other devices which can be determined by turning the equipment off and on the user is encouraged to try to correct the...

Page 15: ...e 39 F Watchdog Fault Turn Power off Turn Power back on If error persists contact MERIT Customer Service 41 53 W RAM Power On Self Test Turn Power off Turn Power back on If error persists contact MERI...