APPENDIX D  CONFORMANCE TO STANDARDS 

The MetaSTAR RF Generator conforms to the following International Standards: 

 

EN 60601-1 

Medical Electrical Equipment. Part 1: General Requirements for 
Basic Safety and Essential Performance 

 

AAMI ES 60601-1 

Medical Electrical Equipment. Part 1: General Requirements for 
Basic Safety and Essential Performance 

 

CSA C22.2 No. 60601-1 

Medical Electrical Equipment. Part 1: General Requirements for 
Basic Safety and Essential Performance 

 

IEC 60601-1-2 

Medical Electrical Equipment, Part 1-2: Collateral  Standard: 
Electromagnetic Compatibility 

 

IEC 60601-2-2 

Medical Electrical Equipment, Part 2-2: Essential Performance 
Particular Requirements for the Basic Safety and Essential Performance 
of High Frequency Surgical Equipment and high frequency surgical 
accessories 

 

ANSI/AAMI/ISO 15223 

Graphical Symbols for Use on Equipment 

 

EN 60529  

Degrees of Protection Provide by Enclosures (IP Code) 

 

ANSI/AAMI ES-1  

Safe Current Limits for Electro medical Apparatus 

 

ASTM D4169 

Standard Practice for Performance Testing of Shipping  
Containers & Systems 

 

EN 55011 (CISPR 11) 

Limits and Methods of Measurement of Electromagnetic Disturbance 
Characteristics of Industrial, Scientific, and Medical (ISM) RF 
Equipment 

 

MDD 93/42/EEC  

Medical Device Directive (Council Directive Concerning Medical 
Devices) 

APPENDIX E  COMPATIBLE POWER CORDS 

Continental Europe, 2.0m, 10A, Inter Power PN 86231000 
Argentina, 2.5m, 10A, Inter Power PN 86270010 
Brazil, 2.5m, 10A, Inter Power PN 86286110 
China, 2.5m, 10A, Inter Power PN 86517040 
India / South Africa, 2.5m, 10A, Inter Power PN 86265010 
Israel, 2.5m, 10A, Inter Power PN 86275010 
Italy, 2.5m, 10A, Inter Power PN 86394000 
Japan, 2.3m, 7A, Inter Power PN 86589000 
Switzerland, 2.5m,10A, Inter Power PN 86285010 
United Kingdom / Ireland, 2.5m,10A, Inter Power PN 86240060 

Hospital-Grade 

Australia/New Zealand, 2.5m, 10A, Inter Power PN 86210140 
Denmark, 2.5m, 10A: Inter Power PN 86280810 
North America, 3.0m, 10A Inter Power 86610710 rev B 

APPENDIX F  TONES 

 

Power On

 

– Speaker: 250Hz for 75ms, 300Hz for 100ms, 400Hz for 100ms, 500Hz for 80ms, 

530Hz for 100ms, 600Hz for 100ms, 750Hz for 100ms, 900Hz for 100ms, 100ms silence, 1000Hz 
110ms five times within 5 seconds,  1000Hz for 500ms tone once at power up. 

 

RF Enabled

 

– Speaker: 300Hz for 200ms – once every two seconds when RF energy is 

delivered in Ablation Mode. 

 

Fault

 - 

High Priority tone. 

 

Warning

 - 

Medium Priority tone. 

 

Button Click

 - The tone pattern shall be the following: 1000Hz for 20ms. 

 

Temperature High

 

– Speaker: 600Hz for 200ms – once every two seconds when RF energy is 

delivered in Ablation Mode and the corresponding TC reading is above 45 °C. 

 

 

Summary of Contents for MetaSTAR 3195

Page 1: ...West Merit Parkway South Jordan Utah 84095 U S A 1 801 253 1600 U S A Customer Service 1 800 356 3748 Authorized Representative Merit Medical Ireland Ltd Parkmore Business Park West Galway Ireland 31...

Page 2: ...PPLIED 4 7 1 Accessories 4 7 2 Single Use Devices See table below for other devices required for Tumor Ablation 4 8 INSTRUCTIONS FOR USE 5 8 1 Set Up Error Bookmark not defined 8 2 MetaSTAR RF Generat...

Page 3: ...FU for the applicable device s used during the procedure Warning Hazardous electrical output This equipment is for use only by qualified medical personnel trained in the use of the MetaSTAR RF Generat...

Page 4: ...istatic sheeting is recommended Caution Cables to the surgical electrodes are recommended to be positioned such that contact with patient or other leads is avoided Caution Do not continue to deliver R...

Page 5: ...SpineSTAR Ablation Instrument Hand Switch Cable andAE Cable are available as backup prior to use 8 3 MetaSTAR RF Generator Nomenclature Symbols ON power Type BF Applied Part Volume Control OFF power W...

Page 6: ...LOW LED 4 RF ON BLUE LED 5 POWER ON GREEN LED 6 Soft Key Button 1 7 Soft Key Button 2 8 Soft Key Button 3 9 Soft Key Button 4 10 Hand Switch Cable Socket 11 AE Cable Socket 12 Fuse Drawer 13 ON OFF Ma...

Page 7: ...be connected to the AE Cable socket on the MetaSTAR RF Generator using the AE Cable inspect the AE Cable for damage If the SpineSTAR Ablation Instrument is connected properly Distal and Proximal tempe...

Page 8: ...TAR Ablation Instrument is properly placed 1 Both thermocouple temperatures are displayed in amber 30 C Note temperatures are displayed in blue when temperature is 30 C 2 Electrode is extended to the...

Page 9: ...instructions for use and maintenance instructions The sole obligation of MERIT under this warranty shall be limited to the replacement of the MetaSTAR RF Generator or equivalent models at no charge t...

Page 10: ...erit Medical Systems Inc 1600 West Merit Parkway South Jordan Utah 84095 U S A 1 801 253 1600 U S A Customer Service 1 800 356 3748 Authorized Representative Merit Medical Ireland Ltd Parkmore Busines...

Page 11: ...operation Enclosure IP 21 III Operating Conditions Temperature 59 F to 104 F 15 C to 40 C Relative Humidity 30 to 75 non condensing Atmospheric Pressure 878 to 1082cmH2O 86 to 106kPa Whenever the tem...

Page 12: ...Instrument are intended for use in the electromagnetic environment specified below The customer or the user of the System should assure that it is used in such an environment Immunity test IEC 60601...

Page 13: ...ntific and Medical ISM RF Equipment MDD 93 42 EEC Medical Device Directive Council Directive Concerning Medical Devices APPENDIX E COMPATIBLE POWER CORDS Continental Europe 2 0m 10A Inter Power PN 862...

Page 14: ...in a particular installation If this equipment does cause harmful interference to other devices which can be determined by turning the equipment off and on the user is encouraged to try to correct the...

Page 15: ...e 39 F Watchdog Fault Turn Power off Turn Power back on If error persists contact MERIT Customer Service 41 53 W RAM Power On Self Test Turn Power off Turn Power back on If error persists contact MERI...

Reviews: