Design and Function
2.11 Ambient conditions
Ź
The incubator may only be used in enclosed rooms and under the following environmental
conditions:
Ambient temperature: 5 ºC to 35 ºC
Humidity:
max. 80 % not condensing
Degree of pollution:
2
Altitude of installation max. 3,000 m above sea level
Ź
The incubator may not be used in areas where there is a risk of explosions. The ambient air
must not contain any explosive dusts, gases, vapours or gas-air mixtures. The incubator is
not explosion-proof.
Ź
Heavy dust production or aggressive vapours in the vicinity of the appliance could lead to
sedimentation in the chamber interior and as a consequence, could result in short circuits
or damage to electrical parts. For this reason, sufficient measures should be taken to
prevent large clouds of dust or aggressive vapours from developing.
2.12 Accessories included
For incubators with basic fittings:
g
Ź
Two (INCO 108) or three (INCO 153 and 246) sliding shelves
Ź
Gas pressure hose
Ź
Water tray (one for INCO 108 and 153 incubators, two for INCO 246 incubators)
Ź
Sterilisation chipcard
Additionally, for models with humidity module:
y,
y
Ź
Water supply tank and connection hose
Additionally, for models with CO
y,
22
or premium module:
p
Ź
Second gas pressure hose with quick-release connector
Additionally, for models with O
y,
22
module:
Ź
Second gas pressure hose with quick-release connector