7.9.1 Description
With the USER ID function, you can lock the settings of individual (e.g. temperature) or
all parameters, so that they cannot be changed at the appliance by accident or unau-
thorised persons. You can also lock setting options in menu mode (e.g. adjustment or
date and time settings) this way.
If adjustment options are locked, this is indicated by the lock symbol in the respective
display.
USER ID data is entered in the AtmoCONTROL software and saved on the USB storage
medium. The USB storage medium is thus acting as a key: Parameters can only be
locked or unlocked if it is connected.
A description of how to create a USER ID in AtmoCONTROL is provided in the separate
AtmoCONTROL manual.
7.9.2 USER ID activation and deactivation
1. Connect the USB storage medium with the USER ID data to the USB port on the
right of the ControlCOCKPIT.
USER-ID
Deactivate
Activate
2. Activate the USER ID. To do so, press the activation key on the right side of the
USER ID display. The display is enlarged and the entry Activate automatically
highlighted.
USER-ID
Deactivate
Activate
3. Con
fi
rm the activation by pressing the con
fi
rmation key. The new USER ID data
are transferred from the USB storage medium and activated. As soon as activa-
tion is complete, a check mark appears in front of the corresponding entry.
4. Remove the USB storage medium. Locked parameters are indicated by the lock
symbol on the respective display.
To unlock the appliance, connect the USB storage medium, activate the USER ID dis-
play and select the entry Deactivate.